dc.contributor.author | Coldur, Fatih | |
dc.contributor.author | Andac, Muberra | |
dc.date.accessioned | 2020-06-21T14:06:28Z | |
dc.date.available | 2020-06-21T14:06:28Z | |
dc.date.issued | 2013 | |
dc.identifier.issn | 1040-0397 | |
dc.identifier.issn | 1521-4109 | |
dc.identifier.uri | https://doi.org/10.1002/elan.201200466 | |
dc.identifier.uri | https://hdl.handle.net/20.500.12712/15964 | |
dc.description | ANDAC, Muberra/0000-0001-7262-9762 | en_US |
dc.description | WOS: 000316213900021 | en_US |
dc.description.abstract | A potentiometric flow injection system was adopted by using a Li+-selective electrode with improved selectivity for sensitive and selective determination of Li+ in human serum spiked samples. For the optimization of the offered flow injection system, multivariate method was utilized. Potentiometric characteristics of the system were investigated (linear range: 5.0x1051.0x101molL1, slope: 56.93mV per decade, detection limit: 3.4x105molL1, sampling-rate: 40 injections per hour). The detection of Li+ in real serum samples was achieved. The obtained results for the spiked serum samples were in good agreement with IC results at 95% confidence level. | en_US |
dc.description.sponsorship | Scientific and Technological Research Council of Turkey (TUBITAK)Turkiye Bilimsel ve Teknolojik Arastirma Kurumu (TUBITAK) [107T-356)]; Ondokuz Mayis UniversityOndokuz Mayis University [OMU-BAP F-448] | en_US |
dc.description.sponsorship | The authors gratefully acknowledge the financial support from The Scientific and Technological Research Council of Turkey (TUBITAK) (Project No: 107T-356) and Ondokuz Mayis University (Project No: OMU-BAP F-448). | en_US |
dc.language.iso | eng | en_US |
dc.publisher | Wiley-V C H Verlag Gmbh | en_US |
dc.relation.isversionof | 10.1002/elan.201200466 | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Lithium determination | en_US |
dc.subject | Flow injection analysis | en_US |
dc.subject | Serum analysis | en_US |
dc.subject | Experimental design | en_US |
dc.subject | Multivariate optimization | en_US |
dc.title | A Flow-Injection Potentiometric System for Selective and Sensitive Determination of Serum Lithium Level | en_US |
dc.type | article | en_US |
dc.contributor.department | OMÜ | en_US |
dc.identifier.volume | 25 | en_US |
dc.identifier.issue | 3 | en_US |
dc.identifier.startpage | 732 | en_US |
dc.identifier.endpage | 740 | en_US |
dc.relation.journal | Electroanalysis | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |