dc.contributor.author | Bilen, CY | |
dc.contributor.author | Asci, R | |
dc.contributor.author | Sarikaya, S | |
dc.contributor.author | Buyukalpelli, R | |
dc.contributor.author | Yilmaz, AF | |
dc.date.accessioned | 2020-06-21T15:43:57Z | |
dc.date.available | 2020-06-21T15:43:57Z | |
dc.date.issued | 2003 | |
dc.identifier.issn | 0892-7790 | |
dc.identifier.uri | https://doi.org/10.1089/089277903769013630 | |
dc.identifier.uri | https://hdl.handle.net/20.500.12712/21745 | |
dc.description | Asci, Ramazan/0000-0002-2119-8963; Asci, Ramazan/0000-0002-2119-8963 | en_US |
dc.description | WOS: 000185358600011 | en_US |
dc.description | PubMed: 14565879 | en_US |
dc.description.abstract | Purpose: We developed a laser pointer device that is mounted on the image intensifier of a C-arm fluoroscope. The laser-assisted fluoroscopic puncture technique (LAFPT) is an alternative method of accessing the kidney. Materials and Methods: The LAFPT was assessed both in an in vitro model and in a series of cases. The in vitro model was composed of Foley balloons filled with contrast medium of different volumes (2-6 mL) embedded in silicone pads. The total fluoroscopy times and attempts required for a successful puncture with LAFPT were compared with those of conventional fluoroscopic access. Results: The number of attempts required to puncture the 2-mL balloons was significantly greater with the conventional fluoroscopic technique than with LAFPT. The fluoroscopy time in vitro was reduced 50% with LAFPT. However, there was no statistically significant difference in the time required for the two techniques in the clinical series. Conclusion: The laser-assisted fluoroscopic puncture technique allows accurate positioning of a needle with reduced radiation exposure. | en_US |
dc.language.iso | eng | en_US |
dc.publisher | Mary Ann Liebert Inc Publ | en_US |
dc.relation.isversionof | 10.1089/089277903769013630 | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.title | Laser-assisted fluoroscopic puncture: A new technique for accessing the kidney | en_US |
dc.type | article | en_US |
dc.contributor.department | OMÜ | en_US |
dc.identifier.volume | 17 | en_US |
dc.identifier.issue | 7 | en_US |
dc.identifier.startpage | 485 | en_US |
dc.identifier.endpage | 491 | en_US |
dc.relation.journal | Journal of Endourology | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |