Effect of Surgical Drill Guide on Heat Generated From Implant Drilling

Date

2009

Author

Misir, A. Ferhat
Sumer, Mahmut
Yenisey, Murat
Ergioglu, Erol

Metadata

Show full item record

Abstract

Purpose: The purpose of this study was to evaluate the heat generated in bone by 2 implant drill systems in vitro with and without using surgical drill guides. Materials and Methods: Temperature was measured with K type thermocouple in vitro using the bovine femoral cortical bone model. A constant drill load of 2.0 kg was applied throughout the drilling procedures via a drilling rig at a speed of 1,500 rpm. Two different implant drill systems-System A (with external irrigation) and System B (with both external and internal irrigation)-were evaluated. The samples were divided into 4 groups. System A test group 1 (TG1) included the following: 20 samples for drilling with surgical drill guides; control group I (CG1): 20 samples for classical implant site preparation. System B test group 2 (TG2) included the following: 20 samples for drilling with surgical drill guides; control group 2 (CG2): 20 samples for classical implant site preparation. Heat was measured at the final drill in the drilling sequence (4.2 and 4.4 mm). Thermocouples were placed at a I-mm distance from the osteotomy area at depths of 3, 6, and 9 mm. Heat measurements were recorded out to 50 uses by a software program. Data were subjected to the Mann-Whitney U and Kruskal-Wallis tests. The a level was set a priori at 0.05. Results: The mean maximum temperatures at the depths of 3, 6, and 9 mm using surgical drill guides were 34.2 degrees, 39.7 degrees, and 39.8 degrees C, respectively, although without using surgical drill guides the values were 28.8 degrees, 30.7 degrees, and 31.1 degrees C. A statistically significant difference was found at the depths of 3, 6, and 9 mm between using surgical drill guides and classical drilling procedure. Conclusions: From a heat generation standpoint, we conclude that preparing an implant site with using surgical drill guides generates heat more than classical implant site preparation regardless of the irrigation type. (C) 2009 American Association of Oral and Maxillofacial Surgeons J Oral Maxillofac Surg 67:2663-2668, 2009

Source

Journal of Oral and Maxillofacial Surgery

Volume

67

Issue

12

URI

https://doi.org/10.1016/j.joms.2009.07.056
https://hdl.handle.net/20.500.12712/18336

Collections

  • PubMed İndeksli Yayınlar Koleksiyonu [6144]
  • Scopus İndeksli Yayınlar Koleksiyonu [14046]
  • WoS İndeksli Yayınlar Koleksiyonu [12971]



DSpace software copyright © 2002-2015  DuraSpace
Contact Us | Send Feedback
Theme by 
@mire NV
 

 




| Policy | Guide | Contact |

DSpace@Ondokuz Mayıs

by OpenAIRE

Advanced Search

sherpa/romeo

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsTypeLanguageDepartmentCategoryPublisherAccess TypeInstitution AuthorThis CollectionBy Issue DateAuthorsTitlesSubjectsTypeLanguageDepartmentCategoryPublisherAccess TypeInstitution Author

My Account

LoginRegister

Statistics

View Google Analytics Statistics

DSpace software copyright © 2002-2015  DuraSpace
Contact Us | Send Feedback
Theme by 
@mire NV
 

 


|| Policy || Library || Ondokuz University || OAI-PMH ||

Ondokuz Mayıs University, Samsun, Turkey
If you find any errors in content, please contact:

Creative Commons License
Ondokuz University Institutional Repository is licensed under a Creative Commons Attribution-NonCommercial-NoDerivs 4.0 Unported License..

DSpace@Ondokuz Mayıs:


DSpace 6.2

tarafından İdeal DSpace hizmetleri çerçevesinde özelleştirilerek kurulmuştur.