Deformation and Fracture of Mtwo Rotary Nickel-Titanium Instruments After Clinical Use

Tarih

2009

Yazar

Inan, Ugur
Gonulol, Nihan

Üst veri

Tüm öğe kaydını göster

Özet

Introduction: In recent years, a number of rotary nickel titanium (NiTi) systems have been developed to provide better, faster, and easier cleaning and shaping of the root canal system. Although the NiTi instruments are more flexible than the stainless steel files, the main problem with the rotary NiTi instruments is the failure of the instruments. The aim of this study was to evaluate the deformation and fracture rate of Mtwo rotary nickel-titanium instruments (VDW, Munich, Germany) discarded after routine clinical use. Methods: A total of 593 Mtwo rotary NiTi instruments were collected after clinical use from the clinic of endodontics over 12 months. The length of the files was measured using a digital caliper to determine any fracture, and then all the files were evaluated under a stereomicroscope for defects such as unwinding, curving, or bending and fracture. The fracture faces of separated files were also evaluated under a scanning electron microscope. The data were analyzed using a chi-square and z test. Results: A percentage of all files (25.80%) showed defects, and the major defect was fracture (16.02%). The most frequently fractured file was #10.04 (30.39%). Deformations without fracture were mostly observed on #15.05 files (25.47%). Conclusions: A higher rate of deformation was observed for #10.04 and #15.05 files. Therefore, these files should be considered as single-use instruments. Because cyclic fatigue was the cause of 71.58% of the instrument fractures, it is also important not to exceed the maximum number of usage recommended by the manufacturer and discard the instruments on a regular basis. (J Endod 2009,35:1396-1399)

Kaynak

Journal of Endodontics

Cilt

35

Sayı

10

Bağlantı

https://doi.org/10.1016/j.joen.2009.06.014
https://hdl.handle.net/20.500.12712/18418

Koleksiyonlar

  • PubMed İndeksli Yayınlar Koleksiyonu [6144]
  • Scopus İndeksli Yayınlar Koleksiyonu [14046]
  • WoS İndeksli Yayınlar Koleksiyonu [12971]



DSpace software copyright © 2002-2015  DuraSpace
İletişim | Geri Bildirim
Theme by 
@mire NV
 

 




| Politika | Rehber | İletişim |

DSpace@Ondokuz Mayıs

by OpenAIRE

Gelişmiş Arama

sherpa/romeo

Göz at

Tüm DSpaceBölümler & KoleksiyonlarTarihe GöreYazara GöreBaşlığa GöreKonuya GöreTüre GöreDile GöreBölüme GöreKategoriye GöreYayıncıya GöreErişim ŞekliKurum Yazarına GöreBu KoleksiyonTarihe GöreYazara GöreBaşlığa GöreKonuya GöreTüre GöreDile GöreBölüme GöreKategoriye GöreYayıncıya GöreErişim ŞekliKurum Yazarına Göre

Hesabım

GirişKayıt

İstatistikler

Google Analitik İstatistiklerini Görüntüle

DSpace software copyright © 2002-2015  DuraSpace
İletişim | Geri Bildirim
Theme by 
@mire NV
 

 


|| Politika || Kütüphane || Ondokuz Mayıs Üniversitesi || OAI-PMH ||

Ondokuz Mayıs Üniversitesi, Samsun, Türkiye
İçerikte herhangi bir hata görürseniz, lütfen bildiriniz:

Creative Commons License
Ondokuz Mayıs Üniversitesi Institutional Repository is licensed under a Creative Commons Attribution-NonCommercial-NoDerivs 4.0 Unported License..

DSpace@Ondokuz Mayıs:


DSpace 6.2

tarafından İdeal DSpace hizmetleri çerçevesinde özelleştirilerek kurulmuştur.